- Test Requirements Zero Power Resistance shall be measured using suitable test equipment having an accuracy of ±0.05% or better. All measurements shall be made in a controlled uniform medium capable of maintaining an accuracy of ±0.01°C. The time response of the equipment used to measure the temperature shall be compatible with that of the thermistor under test.
- Mounting Method Suitable test leads/connections shall be used to connect the thermistor under test to the test equipment. For thermistor beads, flat clips shall be used.
|
|
|